Section 8-7 - Replacing OPIO 8 - 103
8-7-13 Replacement of the Speaker Grill Set-T
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
8-7-13-1 Preparations
1.) Remove the OPIO Top Cover Set-T as described in 8-7-2 "Replacement of the OPIO Top Cover
Set-T" on page 8-89
2.) Remove the UOPIO-T Assy as described in 8-7-10 "Replacement of the UOPIO-T Assy" on page
8-98
3.) Remove the Speaker Set-U as described in 8-7-25 "Replacement of the Speaker Set-U" on page
8-122
8-7-13-2 Removal Procedure
8-7-13-3 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
Table 8-49 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 10 minutes No tools required
Cut resistant gloves
Knee Pad
Step Corresponding graphic
1.
Release 12 snap fits to remove T2 Speaker Grill Left/
Right from the console.