DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
Section 8-7 - Replacing OPIO 8 - 105
8-7-15 Replacement of the T2 OPIO Upper Assy
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
8-7-15-1 Preparations
1.) Remove the OPIO Top Cover Set-T as described in 8-7-2 "Replacement of the OPIO Top Cover
Set-T" on page 8-89
2.) Remove the UOPIO-T Assy as described in 8-7-10 "Replacement of the UOPIO-T Assy" on page
8-98
3.) Remove the T2 OPIO Lower Assy as described in 8-7-14 "Replacement of the T2 OPIO Lower
Assy" on page 8-104
8-7-15-2 Removal Procedure
8-7-15-3 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
Table 8-51 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 10 minutes Philips Screwdriver
Cut resistant gloves
Knee Pad
Step Corresponding graphic
1.
Unscrew 7 screws (red circles) and disconnect the
Lower Keyboard to Touch Panel PWA Power & USB
Cable (blue circle) to remove the T2 OPIO
Upper Assy.