DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 102 Section 8-7 - Replacing OPIO
8-7-12 Replacement of the T2 OPIO Cable Hook
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
8-7-12-1 Removal Procedure
8-7-12-2 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
Table 8-48 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 5 minutes Phillips Screwdriver
Cut resistant gloves
Knee Pad
Step Corresponding graphic
1.
Unscrew 2 screws on the left or right side of the T2 OPIO Cable
Hook to remove from the console.