EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ Totus

GE LOGIQ Totus User Manual

GE LOGIQ Totus
644 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #335 background imageLoading...
Page #335 background image
GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
Section 8-4 - Loading / Reloading / Upgrading the Software 8 - 9
8-4-2 LOGIQ Totus Model Software Configuration and Hardware
8-4-3 Software provided with the LOGIQ Totus
Software is provided on a flash drive with base load image and application software. If patch software
has been included, an additional flash drive may be included with patch software. Verify that the
software is current. Updated software may be available. Check for any available FMIs.
8-4-4 Space management - moving all images
Refer to “Recording Images” in Chapter 9 of the Basic User Manual (appropriate version).
Speak with the personnel at the site to determine which patient images need to be backed up prior to
starting.
8-4-5 Backing up the Patient Archive and System Configurations
“Backup and Restore” in Chapter 9 of the Basic User Manual (appropriate version).
Table 8-3 LOGIQ Totus Software Configuration and Hardware
MODEL NUMBER DESCRIPTION SOFTWARE VERSION
5926013 LOGIQ Totus R4.5.x
5943511 LOGIQ Totus HDU R4.5.x
NOTICE
In order to complete a successful restore of the Patient Database, if needed, the images must be
moved away from LOGIQ Totus before doing backup of the Patient Database. Depending on the
location set-up, either move the images to a remote server or to removable media like DVD or CD discs.
As the images are moved, the database will point to the new location.
If the backup procedure is not completed correctly, the images and database information could be lost.
WARNINGWARNING
Before loading software, CONFIRM WITH THE CUSTOMER OR SITE ADMINISTRATOR
whether the patient archive has been encrypted and if the customer has the password
or the key to unlock the data. Failure to obtain those items will cause permanent DATA
LOSS!
CAUTION
An error, or a power loss, may occur during the software loading. Always back up the patient
images, patient archive, and the system configurations before starting a software load.
In order to complete a successful restore of the patient database, if needed, the images must
be moved away from the LOGIQ Totus before backing up the patient database. Depending on
the location set-up, either move the images to a remote server or to removable media like DVD
or CD discs. If the backup procedure is not completed correctly, the images and database
information could be lost.