DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
Section 8-7 - Replacing OPIO 8 - 95
8-7-7 Replacement of the T2 OPIO Base Rear Cover Assy
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
8-7-7-1 Preparations
1.) Remove the OPIO Rear Tray Set-T as described in 8-11-4 "Replacement of the OPIO Rear Tray
Set-T" on page 8-187
2.) Remove the Probe Holder Set-T as described in 8-7-3 "Replacement of the Probe Holder Set-T" on
page 8-90
8-7-7-2 Removal Procedure
8-7-7-3 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
Table 8-43 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 5 minutes Philips Screwdriver
Cut resistant gloves
Knee Pad
Step Corresponding graphic
1.
Unscrew 5 screws to remove the T2 OPIO Base
Rear Cover Assy from the console.