DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 38 Section 8-4 - Loading / Reloading / Upgrading the Software
8-4-11Full disk encryption and replacement procedures (continued)
To complete a replacement procedure with full disk encryption On:
The system may request the recovery keys through the Numeric Key dialog box. This dialog box may
display up to three times (once for each of the encrypted partitions).
1.) Once you select to load C only or System, the system will check for encryption status and if not
suspended or turned off, it will prompt you to enter the recovery keys. On the Numeric Key dialog
box, enter the recovery key and then select Submit and Exit. This step automatically place the
entered number in Memory.
2.) When the Numeric Key displays Second time, select Mem Recall to recall the recovery key from
memory and then select Submit and Exit.
3.) When the Numeric Key displays the third time, select Mem Recall to recall the recovery key from
memory and then select Submit and Exit. If the keys are accepted, the system will continue to load
the software.
4.) Close with the X.
5.) When the Numeric Key displays again, select Mem Recall to recall the recovery key from memory
or manually enter the recovery key.
6.) Close with the X.