Maintenance
BasicSystemFRURepairs(cont'd.)
FRUDescription
VisualInspectionFunctionalCheckoutProcedures
AcquisitionAssembly
1,2,7,81,2,3,*4,*5,*6,*13
BatteryPack
5
1,2,3,12
PowerSupplyAssembly
7,81,2,3,15
SDCard
61,2,3,4
Plastics
7,8
1,2,3(*15–middleplasticonly)
ACPowerCord
4
1,2,3
Non-listedFRUs
7,8
1,2,3(*15–whenACpowermainsdisturbed)
OptionalSystemFRURepairs
FRUDescription
VisualInspectionFunctionalCheckoutProcedures
InternalModem
7,81,2,3,5,14
KISSPump7,81,2,3
ExternalModem
81,2,3,5,14
LANOption
81,2,3,5
BarcodeReader
7,81,2,3,6
Non-FRURepairs
Description
VisualInspectionFunctionalCheckoutProcedures
Nopartsreplaced
4,5,7,81,2,3,*4,5,6
SoftwareUpdate
n/a
1,2,3,*4,5,6
HardwareUpgrade
7,81,2,3,*4,5,6
AnnualElectricalSafety
Checkout
1,3,4,51,2,3,7,8,9,10,11,12,15,*13,14
VisualInspection
Inspectthefollowingforexcessivewearandanysignsofdamage.
1.Checkforbrokenpatientcable/leadwiresandout-of-dateelectrodes.
a.Verifythatpatientcable/leadwirespassinspection.
b.Verifythattheelectrodespassinspection.
2.Discusselectrodeplacement,skinprep,andpatient-relatedrequirementswith
theECGtechnician.
Verifythatthecustomerisfollowingtheproceduresrecommendedinthe
operator’smanual.
Formoreinformation,referto“PreparingthePatient”intheoperator’smanual
foryoursystem.