ElectromagneticCompatibility
EMCExceptionsDisclosure
TypeExceptionElectromagneticEnvironmentGuidance
Electrostatic
discharge
(ESD)
±4kVairdischargefor
PaperQueuesensors
Floorsshouldbewood,concreteorceramictile.
Ifoorsarecoveredwithsyntheticmaterial,
therelativehumidityshouldbeatleast30%.
Conducted
RF
KISSLeadset:
0.3Vrms
Recommendedseparationdistance:
KISSLeadSet:
d=11.6√P
wherePisthemaximumoutputpowerrating
ofthetransmitterinwatts(W)accordingto
thetransmittermanufactureranddisthe
recommendedseparationdistanceinmeters
(m).
RadiatedRFKISSLeadset:
1.0V/m
Recommendedseparationdistance:
KISSLeadSet:
d=3.5
P80MHzto800MHz
d=7√P800MHzto2.5GHz
wherePisthemaximumoutputpowerrating
ofthetransmitterinwatts(W)accordingto
thetransmittermanufactureranddisthe
recommendedseparationdistanceinmeters
(m).
2028451-183EMAC™1600139