EasyManuals Logo
Home>GE>Medical Equipment>MAC 1600

GE MAC 1600 Service Manual

GE MAC 1600
146 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #104 background imageLoading...
Page #104 background image
Maintenance
ElectricalSafetyChecks(cont'd.)
Step
PolarityCondition
1
UUT
2
-ON
ResultLeakage
Current
Limits
1
ACpowercord
groundprong
toexposed
metalsurface
(groundlug)
N/A
______
Pass/Fail
Lessthan
200mΩ
1
NC=NormalCondition;SFC=SingleFaultCondition;N/A=NotApplicable
2
UUT=UnitUnderTest
UpdatingSoftware
SoftwareupdatesareprovidedonanSDcard.Performasoftwareupdateas
describedinthissection.
1.PowerupthesystemwiththesystemconnectedtoACpower.
2.InserttheSDcardwiththesoftwareupdateintheSDcardslotintherearpanel
asshowninthefollowingillustration.
Thegoldcontactsarefaceup.
3.FromtheMainMenu,pressF5(SystemConguration).
104MAC™16002028451-183E