ElectromagneticCompatibility
(cont'd.)
EmissionsTest
Compliance
ElectromagneticEnvironment-
Guidance
RFemissions(Radiated)
•30MHzto1,000MHzIEC
60601-1-2:2007
•CISPR11:2003/A1:2004/
A2:2006
•EN55011:1998/A1:1999/
A2:2002
•30MHzto6.5GHz
FCC47CFRPart15.33
FCC47CFRPart15.109
Group1
ClassB
RFemissions(Conducted)
•150KHzto30MHz
IEC60601-1-2:2007
•CISPR11:2003/A1:2004/
A2:2006
•EN55011:1998/A1:1999/
A2:2002
•150KHzto30MHz
FCC47CFRPart15.107
Group1
ClassB
Group1use:
ThisdeviceusesRFenergyonlyfor
itsinternalfunction.Therefore,itsRF
emissionsareverylowandarenot
likelytocauseanyinterferencein
nearbyelectronicequipment.
ClassBuse
Thisdeviceissuitableforuse
inallestablishments,including
domesticestablishmentsandthose
directlyconnectedtothepublic
low-voltagepowersupplynetwork
thatsuppliesbuildingsusedfor
domesticpurposes.
HarmonicEmissions
•2nd-40thHarmonic
IEC60601-1-2:2007
•IEC61000-3-2:2005
ClassAThisdeviceissuitableforuse
inallestablishments,including
domesticestablishmentsandthose
directlyconnectedtothepublic
low-voltagepowersupplynetwork
thatsuppliesbuildingsusedfor
domesticpurposes.
Voltageuctuations/
Flickeremissions
IEC60601-1-2:2007
•IEC61000-3-3:1994/A2:2005
Complies
Pass
GuidanceAndManufacturer’sDeclaration-
ElectromagneticImmunity
Thissystemisintendedforuseintheelectromagneticenvironmentspeciedinthe
followingtable.Itistheresponsibilityofthecustomerorusertoensurethatthis
systemisusedinsuchanenvironment.
134MAC™16002028451-183E