ElectromagneticCompatibility
GuidanceAndManufacturer'sDeclaration-Electromagnetic
Immunity
Thissystemisintendedforuseintheelectromagneticenvironmentspeciedinthe
followingtable.Itistheresponsibilityofthecustomerorusertoassurethatthis
systemisusedinsuchanenvironment.
ImmunityTest
Compliance
TestLevel
Compliance
Level
ElectromagneticEnvironment
Guidance
PortableandmobileRF
communicationsequipmentshould
notbeusedclosertoanypartof
the[equipmentorsystem],including
cables,thantherecommended
separationdistancecalculatedfrothe
equationapplicabletothefrequency
ofthetransmitter.
Recommendedseparationdistance
ConductedRF
IEC
60601-1-2:2007
IEC
61000-4-6:
2006
3Vrms
150KHzto
80MHz
@2Hzmod.
Multilink
Leadset:
3.0Vrms
ValueLead
set:
3.0Vrms
KISSLead
set:
0.3Vrms
Recommendedseparationdistance:
MultilinkLeadSet:
d=1.2√P
ValueLeadSet:
d=1.2√P
KISSLeadSet:
d=11.6√P
RadiatedRF
IEC
60601-1-2:2007
IEC
61000-4-3:
2006
3V/m
80MHzto
2.5GHz
@2Hzmod.
Multilink
Leadset:
3.0V/m
ValueLead
set:
3.0V/m
KISSLead
set:
1.0V/m
Recommendedseparationdistance:
MultilinkLeadSet:
d=1.2√P80MHzto800MHz
d=2.3√P800MHzto2.5GHz
ValueLeadSet:
d=1.2√P80MHzto800MHz
d=2.3√P800MHzto2.5GHz
KISSLeadSet:
d=3.5√P80MHzto800MHz
d=7√P800MHzto2.5GHz
136MAC™16002028451-183E

Table of Contents

Other manuals for GE MAC 1600

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 1600 and is the answer not in the manual?

GE MAC 1600 Specifications

General IconGeneral
BrandGE
ModelMAC 1600
CategoryMedical Equipment
LanguageEnglish

Related product manuals