Troubleshooting
YoucanturnHookupAdvisoronoroffintheECGmenu.SelectSystemSetup>
ECG>ECGAcquisition.
Checkfordefectiveordate-expiredelectrodes.
Checkfordefective,broken,ordisconnectedleadwires.
Runthe“”onpagetoensurethatallleadwirespassthenoisetest.
ErrorCodes
Noactionisnecessaryforisolatederroroccurrences.However,iftheunitis
malfunctioningandanyofthefollowingerrormessagesarerepeatingand
unrecoverable,replacetheFRUsintheorderlisted.
AcquisitionErrorCodes
Wheremultiplesolutionsaresuggested,performthesuggestedactionsintheorder
given.Onceasolutionrestoresfunctionality,donotattemptthenextsolutioninthe
list.
AcquisitionErrorCodes
ErrorCodeCauseSolution
AcquisitionError
-4
Invalidcommandsentto
acquisitiondriver
1.Checkthecables.
2.Selfcorrectinginternalerror.
3.Reloadtheapplicationsoftware.
AcquisitionError
-3
Acquisitionbufferoverow
1.Checkthecables.
2.ReplacetheCABLEASSYMAC1600
ACQTOMAINBOARDfoundinthe
CableHarnessKit,(PN2035707).
3.ReplacetheASSYACQUISITIONBOARD
MAC1600(2035705-001).
4.ReplacetheASSYMAC1600
MAINBOARDANDETEMODULE
(2035704-001).
AcquisitionError
-2
Acquisitionmodulenot
connectedordetected
1.Checkthecables.
2.ReplacetheCABLEASSYMAC1600
ACQTOMAINBOARDfoundinthe
CableHarnessKit(PN2035707).
3.ReplacetheASSYACQUISITIONBOARD
MAC1600(2035705-001).
4.ReplacetheASSYMAC1600
MAINBOARDANDETEMODULE
(2035704-001).
2028451-183EMAC™1600
47

Table of Contents

Other manuals for GE MAC 1600

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 1600 and is the answer not in the manual?

GE MAC 1600 Specifications

General IconGeneral
BrandGE
ModelMAC 1600
CategoryMedical Equipment
LanguageEnglish

Related product manuals