Maintenance
Thefollowingwindowopens:
8.IftheWARNINGwindowopens,pressF6(OK),andcontinuewiththesoftware
update.
9.Performtheapplicablecheckoutprocedures.
Referto“FunctionalCheckout”onpage99.
ConditioningTheBatteryPack
Tomaintainthestoragecapacityofthebatterypackinstalledinyourdevice,GE
Healthcarerecommendsthatyouconditionthedevice’sbatterypackonceevery
sixmonthstoresettheelectronicfuelgaugeinsidethebattery.Aconditioncycle
consistsofanuninterrupted“charge-discharge-charge”cycle.
Youcanconditionthebatterypackwhileitisinstalledinadevicethatyouarenot
currentlyusingtorecordtestsonapatient.
1.DisconnecttheACpowerfromthedevice.
2.DisplaytheBATTERYSTATUSwindow:
a.OntheMainMenu,pressF5(SystemConguration).
b.PressF6(More)>F6(More)>F5(ServiceSetup).
c.Whenprompted,entertheServicepasswordandpressF6(OK).
Ifyoudonotknowtheservicepassword,contactGEHealthcareTechnical
Support.
d.SelectSystemDiagnostics.
e.SelectBatteryTest.
TheBATTERYSTATUSwindowopens.
3.AllowthebatterytodischargeuntilBatteryChargeRemaining[%]islessthan
90%.
4.TurnoffthedeviceandreconnectittotheACpower.