3-72 Section 3-8 - Storing and Transporting the Unit
21) Lower the control console (monitor) so that it is pressing down on the two cartons.
22) Place the pre-packed manuals in the compartment under the monitor.
23) Insert the two monitor support foams (item 12) under the front of the monitor, one on each side.
24) From above, slip the anti-static cover (item 4) over the unit.
25) Wrap the excess antistatic sheet (item 2) that was previously placed on the platform base around
the unit and secure it in place with adhesive tape.
26) Assemble the wooden crate, as described in the following section.
3-8-7 Assembling the Wooden Shipping Crate
The Vivid™ 3 ultrasound unit is packed in a wooden shipping crate comprising various sections. These
have rebated joints which are fastened together with 32 Clip-lok
TM
clips that fit into slots. The sections
are as follows:
Case base
Left wall
Right wall
Front wall
Rear wall
Top cover/ramp
To ease the identity of the wooden crate sections, there are red marks on the inner surface of the base,
the left wall and rear wall. When the case parts are joined correctly, the three red marks should appear
next to each other in one corner. The top cover has also a red mark that should be facing downwards
in the corner, where the other three marks meet.
Note: Always insert the short leg of a clip in the appropriate slot first, and then, using the heel of your
hand, insert the long leg of the clip into the slot located on the part being joined.
1) Place the right wall (item 6) vertically along the right side of the crate base and secure it to the base
using three clips (item 15).
2) Place the left wall (item 7) vertically along the left side of the case base and secure it to the case
base using three clips (item 15).
3) Place the wooden cover (item 3) on top of the right and left side walls and fasten it with three clips
(item 15) to each of the side walls.
4) Insert the front support foam (item 8) in front of the front wheels.
5) Place the back support foam (item 9) in front of the rear wheels.
6) Place the keyboard support foam (item 10) in front of the keyboard.
7) Place the monitor support back foam (item 11) at the monitor's back.
8) Put the cases containing the probes (item D) and Peripheral options in their original packing
(item E) in the appropriate spaces.
9) Hold the rear wall section (item14) vertically against the rear of the crate and secure it in place using
eight clips (item 15).
NOTICE
Important!
Verify that the height from the crate base to the top of the monitor does not exceed 132cm (52in)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for GE Vivid 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid 3 and is the answer not in the manual?

GE Vivid 3 Specifications

General IconGeneral
BrandGE
ModelVivid 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals