2-2-6 EMI Limitations
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. They also generate EMI. The Vivid™ 3 ultrasound
unit complies with limits as stated on the EMC label. However, there is no guarantee that interference
will not occur in a particular installation.
Note: Possible EMI sources should be identified before the unit is installed, and should not be on the
same line as the ultrasound system. A dedicated line should be used for the ultrasound system.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. Sources
of EMI include the following:
Medical lasers.
Scanners.
Cauterizing guns.
Computers.
Monitors.
Fans.
Gel warmers.
Microwave ovens.
Portable phones.
Broadcast stations and mobile broadcasting machines.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for GE Vivid 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid 3 and is the answer not in the manual?

GE Vivid 3 Specifications

General IconGeneral
BrandGE
ModelVivid 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals