GE MEDICAL SYSTEMS
D
IRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
6-2 Section 6-2 - Input AC Voltage Configuration
Section 6-2
Input AC Voltage Configuration
The internal Vivid™ 3 voltage is 220V AC.
The Back End and Front End Crates operate on 220V AC.
The Monitor and Peripherals operate on local voltage.
6-2-1 Secondary Voltage Configuration
The AC input voltage is factory preset according to specific requirements, as indicated on the label near
the AC input connector. The AC transformer’s primary and secondary voltages are set up by using the
appropriate plugs connected in Jumper1 and Jumper2 on the AC Distribution Box, in accordance with
the information in
Table 6-2 below:
6-2-2 AC Input Cord
The input voltage power cord and plug must meet the local standards and requirements.
Table 6-2 Primary and Secondary AC Voltages
Jumper 1
PRIMARY
Voltage
Range
100-120V 220-240V
Plug P/N
2253054 2253056
Jumper 2
SECONDARY
Voltage
100V 120V 220-230V 230-240V
Plug P/N
2253057 2253058 2253059-2 2253060-2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for GE Vivid 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid 3 and is the answer not in the manual?

GE Vivid 3 Specifications

General IconGeneral
BrandGE
ModelVivid 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals