dmm.opendetector open voltage
With dmm.opendetector = dmm.ON, a separate -1.5 A I
OPENLEAD
SHI and a separate SLO current
source will pulse on and off before the start of each measurement while I
SOUR
remains enabled. The
A/D will monitor SHI for 2 ms, then switch to SLO for an additional 2 ms. During either phase, if the
input voltage exceeds the above table, the A/D will stop in less than 100 µs and return an overflow
reading. If there are no open leads detected during the I
OPENLEAD
phase, the I
OPENLEAD
is disabled and
standard 4-wire is enabled.
V
MEAS
with open input:
If Sense HI is disconnected, V
MEAS
will droop less than –1V, causing an A/D overflow.
V
MEAS
with valid connections:
For valid connections, INPUT Sense HI, V
MEAS
, will dip during the 4 ms I
OPENLEAD
phase. The amount
of the voltage dip is the sum of I
OPENLEAD
and the range I
SOUR
and R
DUT
load. For example, if
measuring a 100 kÎ©ï€ on the 100 kÎ©ï€ range, the V
MEAS
across the 100 kΩ will be 0.85 V (10 A to
1.5 µA) × 100 KÎ©ï€ during I
OPENLEAD
and 1 V during measurement phase.
The tables below note timing with dmm.opendetector = dmm.ON.