Section 8: Theory of operation Series 3700A System Switch/Multimeter Reference Manual
8-16 3700AS-901-01 Rev. D/June 2018
dmm.opendetector open voltage
With dmm.opendetector = dmm.ON, a separate -1.5 A I
OPENLEAD
SHI and a separate SLO current
source will pulse on and off before the start of each measurement while I
SOUR
remains enabled. The
A/D will monitor SHI for 2 ms, then switch to SLO for an additional 2 ms. During either phase, if the
input voltage exceeds the above table, the A/D will stop in less than 100 µs and return an overflow
reading. If there are no open leads detected during the I
OPENLEAD
phase, the I
OPENLEAD
is disabled and
standard 4-wire is enabled.
V
MEAS
with open input:
If Sense HI is disconnected, V
MEAS
will droop less than –1V, causing an A/D overflow.
V
MEAS
with valid connections:
For valid connections, INPUT Sense HI, V
MEAS
, will dip during the 4 ms I
OPENLEAD
phase. The amount
of the voltage dip is the sum of I
OPENLEAD
and the range I
SOUR
and R
DUT
load. For example, if
measuring a 100 kΩon the 100 kΩrange, the V
MEAS
across the 100 kΩ will be 0.85 V (10 A to
1.5 µA) × 100 KΩduring I
OPENLEAD
and 1 V during measurement phase.
The tables below note timing with dmm.opendetector = dmm.ON.
Range
SHI and SLO
IOPENLEAD
phase (ms)
1
SHI settling
time (ms)
Line frequency
(Hz)
SHI measurement time
(ms)
1-10 kΩ
4.0
0.5
60
min
max
0.0083
250
50
0.010
240
100 kΩ
4.0
2.0
60
0.0083
250
50
0.010
240
1 MΩ
4.0
30.0
60
0.0083
250
50
0.010
240
10 MΩ to
100 MΩ
4.0
5.0
1
60
0.0083
250
Range
Internal
DMM comm
(ms)
SLO
settling
time (ms)
SLO
measurement
time (ms)
Internal
DMM comm
1 kΩ to
10 kΩ
0.06
0.5
min
max
0.06
0.0083
250
0.010
240
100 kΩ
0.06
1.0
0.0083
250
0.06
0.010
240
1 MΩ
0.06
1.0
0.0083
250
0.06
0.010
240

Table of Contents

Other manuals for Keithley 3721-ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3721-ST and is the answer not in the manual?

Keithley 3721-ST Specifications

General IconGeneral
BrandKeithley
Model3721-ST
CategoryMultimeter
LanguageEnglish

Related product manuals