Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-334 3700AS-901-01 Rev. D/June 2018
scan.trigger.measure.stimulus
This attribute selects the trigger stimulus of the event detector trigger.
Instrument reset
Recall setup
Scan reset
Create
configuration script
Save setup
Usage
eventID = scan.trigger.measure.stimulus
scan.trigger.measure.stimulus = eventID
The trigger stimulus that is used for the measurement event
Details
This attribute selects the events that cause the measurement event detector to enter the detected
state. You can use this to pace each one of the measurement count readings with an event.
To pace all readings by a single event, use scan.trigger.sequence.stimulus.
To bypass waiting for an event, set this to 0.
eventID can be set to one of the existing trigger event IDs, shown in the following table.
channel.trigger[N].EVENT_ID or 41 to 48
The trigger event generated by the channel trigger N.
digio.trigger[N].EVENT_ID or 1 to 14
An edge (either rising, falling, or either based on the
configuration of the line) on the digital input line.
display.trigger.EVENT_ID or 39
The trigger key (TRIG) on the front panel is pressed.
dmm.trigger.EVENT_LIMIT1_HIGH or 53
A DMM trigger event that indicates a measurement has
exceeded the high limit value on limit 1.
dmm.trigger.EVENT_LIMIT1_LOW or 52
A DMM trigger event that indicates a measurement has
exceeded the low limit value on limit 1.
dmm.trigger.EVENT_LIMIT2_HIGH or 55
A DMM trigger event that indicates a measurement has
exceeded the high limit value on limit 2.
dmm.trigger.EVENT_LIMIT2_LOW or 54
A DMM trigger event that indicates a measurement has
exceeded the low limit value on limit 2.
A *trg message on the active command interface. If
GPIB is the active command interface, a GET message
also generates this event.
trigger.blender[N].EVENT_ID or 58 to 59
A combination of events has occurred.
trigger.timer[N].EVENT_ID or 20 to 23
tsplink.trigger[N].EVENT_ID or 17 to 19
An edge (either rising, falling, or either based on the
configuration of the line) on the TSP-Link trigger line.
lan.trigger[N].EVENT_ID or 29 to 36
A LAN trigger event has occurred.