Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-104 3700AS-901-01 Rev. D/June 2018
channel.trigger[N].wait()
This function waits for the specified trigger or timeout period, whichever comes first.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
triggered = channel.trigger[N].wait(timeout)
triggered
Returns an indication that a trigger occurred
N
Number indicating the channel trigger to wait for (1 to 8)
timeout
The number of seconds to wait
Details
If one or more trigger events were detected since the last time channel.trigger[N].wait or
channel.trigger[N].clear was called, this function returns immediately.
After waiting for a trigger with this function, the event detector is automatically reset and rearmed.
This is true regardless of the number of events detected.
The value for timeout must be greater than zero and less than 10,000.
Example
channel.trigger[1].wait(5)
Wait 5 seconds for channel trigger event 1 to
occur or timeout if trigger event is not detected in
5 seconds.
Also see
None

Table of Contents

Other manuals for Keithley 3721-ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3721-ST and is the answer not in the manual?

Keithley 3721-ST Specifications

General IconGeneral
BrandKeithley
Model3721-ST
CategoryMultimeter
LanguageEnglish

Related product manuals