Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-101
channel.trigger[N].clear()
This function clears any pending events.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
channel.trigger[N].clear()
N
Number indicating the trigger line to clear (1 to 8)
Details
This function clears any pending events for the channel trigger specified by N.
Example
channel.trigger[1].clear()
Clears any pending events on channel
trigger 1.
Also see
channel.trigger[N].set() (on page 11-103)
channel.trigger[N].EVENT_ID
This constant indicates the trigger event generated by the channel trigger N.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
X = channel.trigger[N].EVENT_ID
X
The trigger event number
N
Number indicating the channel trigger event ID (1 to 8)
Example
scan.trigger.channel.stimulus = channel.trigger[1].EVENT_ID
Use channel trigger 1 events to pace the channel action of the scanning or set the trigger stimulus of the
channel event detector to channel trigger 1.
Also see
channel.trigger[N].set() (on page 11-103)

Table of Contents

Other manuals for Keithley 3721-ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3721-ST and is the answer not in the manual?

Keithley 3721-ST Specifications

General IconGeneral
BrandKeithley
Model3721-ST
CategoryMultimeter
LanguageEnglish

Related product manuals