Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-50 3700AS-901-01 Rev. D/June 2018
channel.connectsequential
This attribute controls whether or not channels are closed sequentially.
Instrument reset
Recall setup
Create configuration script
Usage
sequential = channel.connectsequential
channel.connectsequential = sequential
â–ª channel.OFF or 0: Disable sequential connections
â–ª channel.ON or 1: Enable sequential connections
Details
When channel.connectsequential is enabled, the list of channel actions is acted on
sequentially. No two relays are opened or closed simultaneously.
Using a sequential close allows you to determine the time for a close operation to happen. For
example, if you close three channels and each takes 4 ms to close (assuming no additional user
delay times), with sequential on, it will take 12 ms. With sequential off, it may be 4, 8 or 12 ms,
depending on whether or not the card can close multiple channels at once.
The order in which channels are opened or closed is not guaranteed with sequential off.
The sequential setting affects all channels in the instrument.
Example
channel.connectsequential = channel.ON
Specifies that channels close sequentially.
Also see
channel.connectrule (on page 11-48)
Switching and scanning (on page 5-1)