Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-407
trigger.blender[N].orenable
This attribute selects whether the blender performs OR operations or AND operations.
Instrument reset
Trigger blender N reset
Recall setup
Create configuration script
Save setup
Usage
orenable = trigger.blender[N].orenable
trigger.blender[N].orenable = orenable
The type of operation:
â–ª true: OR operation
â–ª false: AND operation
The blender number (up to two)
Details
This command selects whether the blender waits for any one event (OR) or waits for all selected
events (AND) before signaling an output event.
Example
trigger.blender[1].orenable = true
trigger.blender[1].stimulus[1] = digio.trigger[3].EVENT_ID
trigger.blender[1].stimulus[2] = digio.trigger[5].EVENT_ID
Generate a trigger blender 1
event when a digital I/O
trigger happens on line 3 or
5.
Also see
trigger.blender[N].reset() (on page 11-408)
trigger.blender[N].overrun
This attribute indicates whether or not an event was ignored because of the event detector state.
Instrument reset
Trigger blender N clear
Trigger blender N reset
Usage
overrun = trigger.blender[N].overrun
Trigger blender overrun state (true or false)
The blender number (up to two)