Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-318 3700AS-901-01 Rev. D/June 2018
scan.list()
This function queries the active scan list.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Instrument reset
Channel reset
Scan reset
Recall setup
Change of channel or scan setting
Create configuration script
Save setup
Empty list
Usage
scanList = scan.list()
scanList
A string that lists the existing scan step information
Details
This function lists the existing scan list.
When you change a channel or scan attribute for an existing scan list item, the scan list is recreated
based on this change. If the scan list cannot be rebuilt, the scan list is lost.
To avoid unintentional changes to an existing scan list, configure channel and scan settings before
adding or creating a scan or step using the commands scan.add(), scan.addimagestep(), and
scan.create().
If the scan list is empty, the string "Empty Scan" is returned. Otherwise, the string lists each step in
the scan along with its information for step, open, measure configuration, count, and close (see the
example below).

Table of Contents

Other manuals for Keithley 3721-ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3721-ST and is the answer not in the manual?

Keithley 3721-ST Specifications

General IconGeneral
BrandKeithley
Model3721-ST
CategoryMultimeter
LanguageEnglish

Related product manuals