AS3000™ Service Manual 0070-10-0683 6 - 23
Periodic Maintenance Functional Tests
3. Proceed based on one of the following two conditions:
If the calibration is successful, the screen shown in FIGURE 6-25 will be displayed,
instructing the user to reinstall the oxygen sensor into the Breathing System. Select the
Done button to complete the process.
If the calibration fails, the screen shown in FIGURE 6-26 will be displayed,
instructing the user to either repeat the calibration (by selecting the Repeat Cal button)
or to replace the oxygen sensor. If the oxygen sensor must be replaced, select the Exit
button, replace the oxygen sensor using a push and turn motion, and then repeat the
calibration.
FIGURE 6-23 Oxygen Sensor
Calibration Instructions
FIGURE 6-24 Oxygen Sensor Calibration
Progress Bar
FIGURE 6-25 Oxygen Sensor
Calibration Successful
FIGURE 6-26 Oxygen Sensor
Calibration Failed