7-44 Structure and Assembly/Disassembly
Fig 7-47 Disassemble the USB adapter board
7.4.11 DVD Assembly
Tool: cross-headed screwdriver. Please refer to Chapter 7.3.1 Tools Required for tools’
specifications.
1. Remove the IO module and the rear cover (refer to chapter 7.4.2 IO Rear Board and
7.4.3 Machine Fan).
2. Remove the M4X8 screws (2 pcs) for fixing the DVD assembly, pull out the two cable
plugs connecting to the main unit, cut off the cable tie to pull DVD assembly
backwards.
Fig 7-48 Disassemble the DVD-R/W assembly
7.4.12 Others
Tools: cross-headed screwdriver, flat-headed screw, inner hexagon spanner, spanner,
diagonal cutting pliers. Please refer to Chapter 7.3.1 Tools Required for tools’
specifications.
7.4.12.1 Spring assembly
1. Remove the control panel assembly (refer to step 1 to 3 in Chapter 7.4.10 Control
Panel Assembly).
2. Remove the IO assembly (refer to step 1 in chapter 7.4.2 IO Rear Board).