9-8 System Diagnosis and Support
Load printing mark
Construct low power consumption
Construct the USB management and
writing management
Initialize video replay device
Initializing ultrasound
image…
Set the virtual machine device, and
initialize the virtual machine
Construct object trees of the front-end
and the back-end (ultrasys and so on)
In increments of 2.
Initializing ultrasound
application…
Add function package of measure menu
Register the application interface of
measure patient
In increments of 1.
Initialization
completed
The architecture of patient information
management (UPatientApp)
In increments of 1.
Hide boot-trap graphic
9.3 Alarming and Abnormal Information
The system has also alarming function, when the machine fails, it would pop up the
alarming dialogue box, and simultaneously generate LOG file saved in the system log,
whose place of saving in HDD is D:/DCN2PLus/Log/PeriLog. The detailed description of
alarming information is as following:
9.3.1 Turning on the System Configuration File is
Abnormal
9.3.2 The voltage of system power is abnormal
Alarming tips
LOG recorder
Suggestion
Fail to open the file "SystemConfiguration.ini",
and please check HDD data.
none
Reinstall the system
software.
Alarming tips
LOG record
Suggestion
The real-time
battery will be
out of power,
please replace
quickly!
×××: System Monitor: Power supply alert! [VBAT],
Current voltage: [VVV] V, Limit voltage: [LLL]-[HHH]
V.
×××represents time, e.g. 2010-12-25 14:15:25
[VVV] represents the current value, and [LLL]-[HHH]
represents the lower and upper limits.
Replace the button
battery.
none
×××: System Monitor: Power supply alert! [XXX],
Current voltage: [VVV] V, Limit voltage:[LLL]-[HHH]V.
×××represents time, e.g. 2010-12-25 14:15:25
[XXX] represents voltage name, [VVV] represents
Short-circuited,
protection and
overvoltage, check
the related circuit.