9-10 System Diagnosis and Support
9.3.5 Battery Alarming
×××represents time, e.g. 2010-12-25 14:15:25
[XXX] Represents fan name, [VVV] represents the
current value, and [LLL] represents the limit. The
detailed fan names are as following:
Main Board Fan (Main unit box fan)、Power Fan
(Machine fan)
again/remove
the barrier
which causes
the fan
malfunction
Alarming! Illegal operating
battery! Lead to permanent
damage
Stop the illegal operation.
Battery communication is
abnormal,
Check the battery connection
or replace the battery
Battery communication is
abnormal, please connect to AC
power supply or shut down!
Otherwise the system shuts
down automatically after 60s!
Battery I2C error, the
battery supply
Check the battery connection
or replace the battery
Battery temperature is too high;
please connect to AC power
supply or shut down! Otherwise
the system shuts down
automatically after 60s!!
Battery temperature is
out of range temp,
Battery temperature is:
xx centigrade
Check if the heat radiating
condition is normal.
Battery voltage is too low;
please connect to AC power
supply or shut down!
Otherwise the system shuts
down automatically after 60s!
battery voltage is low,
battery volt is xx V
Check the battery connection
or replace the battery
Battery error! Battery cannot be
used!
Battery voltage is too low,
please connect to AC power
supply or shut down!
Otherwise the system shuts
down automatically after 60s!
Low battery power,
battery power is XX
Input AC, check if the battery
charging is normal, and it can
be used when full of power.
Battery life is approaching,
please replace with new battery
battery cycle is XX,
please change the
battery
Battery I2C error,
Shutdown State
Replace the battery and
confirm the malfunction
module.
Battery Hot Plug,
Shutdown State
Stop the illegal operation.