10-2 Care and Maintenance
10.1.2 Care and Maintenance Frequency
Table 10-3 Maintenance Frequency
Maintain content
Frequency
Method
Transducer cleaning
Every time after using, 1
time/day
Refer to 10.2.2
Cleaning the cover and control
panel
1 time/month
Same as the above
Cleaning LCD
1 time/month
Same as the above
Cleaning the trackball
1 time/month
Same as the above
Cleaning peripherals and
accessories
1 time/month
Refer to 10.2.2
Earth resistance
1 time per two years
Refer to 10.4.1
Check earth leakage current
1 time per two years
Same as the above
Check patient leakage current
1 time per two years
Same as the above
Mechanical safety checking
1 time/year
Refer to 10.4.2
Measurement accuracy
2 times/year
Refer to 10.3.1
Probe imaging checking
4 times/year
Same as the above
Peripherals and accessories
testing
2 times/year
Refer to 10.3.2
Battery checking
2 times/year
Same as the above