EasyManuals Logo
Home>Mindray>Diagnostic Equipment>ME8

Mindray ME8 Service Manual

Mindray ME8
292 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #277 background imageLoading...
Page #277 background image
Part III Appendix
Electrical Safety Maintenance 260
Electrical safety check
2- Equipment shell and accessories
Test process
Visual inspection
Shell and accessories
No physical damage
The switch, connector, and dashboard are free from physical
damages.
There are no overflowing liquid residues, such as water,
coffee, and chemicals.
The probe is free from physical damages, such as cracks or
swelling on the probe head.
No parts such as buttons, knobs and terminals are loosening
or missing.
Associated check
Shell and accessories
No abnormal sound, such as cracks inside the machine
No abnormal swells, such as the smell of scorch or smoke
from the air vent
No voice prompts regarding equipment bugs or precautions
for the operator
3-Equipment labels
Test process
Check whether labels provided by the equipment vendor are complete and legible.
Labels on the main unit
General warning labels
Slope and HV caution labels
No-press labels
Note:
The 4-protection ground impedance test item is applicable only to the MT3 trolley (with power) and not to the
standalone MX portable device or the MT2 trolley (without power).

Table of Contents

Other manuals for Mindray ME8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray ME8 and is the answer not in the manual?

Mindray ME8 Specifications

General IconGeneral
BrandMindray
ModelME8
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals