Reference Manual V1.19 Software Versions 4.xx
Page 26 003R-682-119
the factory. Each time a critical step is altered, the counter(s) will increase by one. Whenever the
instrument is powered up, or setup mode is entered/exited, the current value in the counter(s) is
displayed briefly (e.g. C00010).
Industrial
OIML
NTEP
The Calibration Counter
increments when trade
critical settings, marked
with , are changed. An
example of the counter is
C.00019.
The Calibration Counter
increments when trade
critical settings, marked
with , are changed. An
example of the counter is
C.00019
The Calibration Counter
increments when trade critical
settings in the Calibration (CAL)
menu, marked with , are
changed. An example of the
counter is C.00010.
The Configuration Counter
increments when other trade
critical settings (i.e. not in the
CAL menu), marked with , are
changed. An example of the
counter is F.00009.
The value(s) of the counter(s) is written on the tamperproof trade label on the front or top of the
indicator for trade-certified applications and functions as an electronic seal. If any legal for trade
settings are changed on the instrument, the current value of the calibration counter(s) will be
different from the recorded value and the seal is broken. In this manual, items marked with
indicate that the setting is legal for trade critical settings.
6.6. Passcodes
The instrument has two levels of passcodes to provide a security lock on accessing Setup via
the keypad.
Full Setup Passcode
Safe Setup Passcode
The Full Setup passcode can also be used to access Safe Setup.
6.6.1. Full Setup Passcode
Setting a passcode for Full Setup restricts any access to Full Setup. Refer to FULL.PC
(Full Security Passcode for Digital Setup) page 36.
6.6.2. Safe Setup Passcode
Setting a passcode for Safe Setup restricts access to Safe Setup functions. Refer to
SAFE.PC
(Safe Security Passcode for Digital Setup) page 36.
6.6.3. Setup Lock-Out
If an attempt is made to enter Full or Safe Setup using an incorrect passcode, the instrument
will respond with the message ENTRY DENIED and then the user will be returned to normal
operating mode. A passcode counter has been set so that only three failed attempts can be
made to access Full/Safe Setup. On the fourth attempt the user will be 'locked out' of
Full/Safe setup. Should this occur the ENTER PASS prompt will not display, but instead the
ENTRY DENIED message displays and returns the user to the normal operating mode. To
rectify this issue the instrument must be turned off. When the instrument is turned back on the
passcode counter is reset to zero (allowing the user to enter the correct passcode).

Table of Contents

Other manuals for Rinstrum R320

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rinstrum R320 and is the answer not in the manual?

Rinstrum R320 Specifications

General IconGeneral
BrandRinstrum
ModelR320
CategoryAccessories
LanguageEnglish

Related product manuals