Troubleshooting and Repair Welch Allyn Monitor Service Utility
Vital Signs Monitor Series 300 41 Service Manual
Characterizing NIBP
To characterize the NIBP module, start the Welch Allyn Monitor Service Utility and click on
Characterize NIBP.
Click Help in the utility menu, and perform the procedure as indicated.
Warning If you cannot characterize the NIBP module, return the monitor to
Welch Allyn for service. (See Returning Products on page 12.)
Do not use the monitor if you cannot characterize the NIBP module. If the
NIBP module is not properly characterized, the monitor could overinflate a
neonatal cuff, which could create a hazard for neonatal patients.