YL9110 Quaternary Pump
Pulsation by compressibility and delay Compensation of pulsation
[Fig. 2] Delivery mechanism of YL9110 quaternary pump
2-4. Solenoid valve control
YL9110 quaternary pump controls open/close signal of solenoid valve to perform low pressure gradient
elution. For accuracy and precision of solvent mixing, YL9110 quaternary pump controls valve opening
and closing time with real time compensation according to time gap of valve cycling and the phase speed
of cam.
[Fig. 3] Compensation of solenoid valve time gap