YL9170 RI Detector
4-2. Lamp exchange and adjustment of the lamp
It might be possible that you need to re-adjust the light source due to some changes in the used light bulb
due to the transport or you need to replace a burned out lamp. The adjustment of the light source is only
possible when the housing is opened. To prepare for adjustment do the following steps :
- Switch off the instrument
- Unplug the main cable to prevent electrical shock
NOTE
Some operations during adjustment have to be done when the instrument is in operation. So you
need to reconnect the main cable and to switch on the YL9170 RI detector. When the instrument is
operated with opened housing make sure that you DO NOT TOUCH any electrical component!
1) Open the outer housing of the optical bench
2) Open the inner black housing of the optical bench.
Optical bench of
YL9170 RI Detector
Y
L9170
[Interior and optical bench of RI Detector]
11