YL9111 Binary Pump
1-1. Specifications
1) Operating principle : X2 Parallel dual-plunger pump, high pressure gradient
2) Compressibility compensation : Automatic
3) Flow range : 0.001-10ml/min
4) Flow rate accuracy : ≤±1% at 1ml/min
5) Flow rate precision : 0.1% RSD at 1ml/min
6) Maximum pressure : 6000 psi
A. Operating range : 0-6000 psi up to 5ml/min
B. Operating range : 0-3000 psi up to 10ml/min
7) Pressure pulsation : ≤±1% at 1ml/min
8) Composition Precision : <0.1%
9) Composition Accuracy : <0.5%
10) Auto prime/purge
11) Communications : LAN
12) Safety & maintenance : Leak detection, Diagnostics, Error detection
13) Dimensions : 385 X 160 X 565mm (width X height X depth)
14) Line Voltage : 100-240VAC, ±10%, automatic voltage selection
15) Line frequency : 50/60Hz, ±5%
16) Power consumption : 140W
[Optional specification]
Head type Narrow Bore Analytical Semi prep
Head volume(μl/each)
25 64 144
Flow range(ml/min) 0.001~5 0.001~10 0.01~50
Flow range(ml/min) 0.005~5 0.01~10 0.05~25
Pressure(psi) 5000psi 6000psi 3500psi
In-line filter
(micron)
2 2 10
Material
Zirconium, Ruby,
PEEK
UHMWPE, PTFE
SS316, Zirconium,
PTFE, Sapphire,
UHMWPE
SUS316, Sapphire,
Ruby, UHMWPE, PTFE