77
Chapter 4 Calibration Procedures
Output Impedance Adjustment
4
4
2 Use the DMM to make a resistance measurement at the front panel
Output connector for each setup in the following table. The expected
measured value is approximately 50.
* Constants are stored after completing this setup.
3 Using the numeric keypad or knob, adjust the displayed impedance at
each setup to match the measured impedance. Select ENTER VALUE.
4 There are no specific operational verification tests for Output Impedance.
Continue with the next adjustment procedure in this chapter.
Setup
8* -30 dB range with distortion filter
9* -20 dB range with distortion filter
10* -10 dB range with distortion filter
11* 0 dB range with distortion filter
12* +10 dB range with distortion filter
13* -30 dB range without distortion filter
14* -20 dB range without distortion filter
15* -10 dB range without distortion filter
16* 0 dB range without distortion filter
17* +10 dB range without distortion filter

Table of Contents

Other manuals for Agilent Technologies 33250A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33250A and is the answer not in the manual?

Agilent Technologies 33250A Specifications

General IconGeneral
BrandAgilent Technologies
Model33250A
CategoryInverter
LanguageEnglish

Related product manuals