91
Chapter 4 Calibration Procedures
Calibration Errors
4
4
Calibration Errors
The following errors are failures that may occur during a calibration.
System error messages are described in chapter 5 of the Agilent 33250A
User’s Guide. Self-test error messages are described in this manual
beginning on page 137.
701 Calibration error; security defeated by hardware jumper
If you short the calibration secure jumper while turning ON the
instrument, this error will occur indicating the security password has
been overwritten.
702 Calibration error; calibration memory is secured
703 Calibration error; calibration memory is secured
704 Calibration error; secure code provided was invalid
705 Calibration error; calibration aborted
706 Calibration error; value out of range
You have typed in a value that was unexpected by the calibration firmware.
For example, if a number is expected such a 50.XX ohms, and you enter
10 ohms, that number is outside the expected range of valid inputs.
707 Calibration error; value out of range
Occurs during the ADC Adjustment, setup 6, if the 1 Volt input voltage is
too high.
850 Calibration error; set up is invalid
You have selected an invalid calibration setup number.
851 Calibration error; set up is out of order
Certain calibration steps require a specific beginning and ending sequence.
You may not enter into the middle of a sequence of calibration steps.