314 Chapter 10
CDMA Digital Modulation
cdma2000 Reverse Link Modulation for Component Test
4. Highlight the Power value (–17.36) in table row 3.
5. Press
Edit Item > 10 > dB.
The display shows that the total power is now at 0.34 dB. You can rescale the total channel power to
0 dB by pressing
Adjust Code Domain Power > Scale to 0 dB.
6. Highlight the Data value (RANDOM) in table row 3.
7. Press
Edit Item > 00110011 > Enter.
The forward reverse channel parameters have now been modified, as shown in Figure 10-5.
Figure 10-5
8. Press
Return.
The text area displays RVS CDMA2000 Setup: SR1 5 Channel (Modified) as the current
configuration. You now have a modified traffic channel with a data rate of 4800 and a power level of 10.00
dB transmitting 00110011.
To store a custom cdma2000 state, see “Storing a Component Test Waveform to Memory” on page 316.
Inserting Additional cdma2000 Reverse Link Traffic Channels
1. Press
Edit Channel Setup.
2. Move cursor to the bottom row and press
Insert Row > Supplemental2 Traffic.
3. Press
Done.

Table of Contents

Other manuals for Agilent Technologies E4428C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4428C and is the answer not in the manual?

Agilent Technologies E4428C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4428C
CategoryInverter
LanguageEnglish

Related product manuals