7Performance Tests
Agilent U1602B/U1604B User’s and Service Guide 133
List of Test Equipment
For scope performance verification, the recommended test equipments are
listed as below.
Table 7 - 1 List of test equipments for scope performance test
Application Equipment Critical Specifications Recommended Equipment
Voltage measurement
accuracy
Calibrator / Power supply 5 mV to 35 VDC, 0.1 mV resolution FLUKE 5520A
Digital multimeter Better than 0.01% accuracy Agilent 34401A
Cable BNC, quantity of 2 Agilent 10503A
Shorting capacitor BNC Agilent 1250-0774
Adapter BNC (f) to banana (m) Agilent 1251-2277
Adapter BNC tee (m) (f) (f) Agilent 1250-0781
Bandwidth Calibrator NA FLUKE 5520A
Feedthrough 50&, BNC connector (m) and (f) Agilent 11048C
Cable Type N (m), 24 inches Agilent 11500B
Horizontal time base Calibrator 5 ppm after 1/2 an hour FLUKE 5520A
accuracy Feedthrough 50&, BNC connector (m) and (f) Agilent 10503A
Cable BNC, 3 feet length Agilent 11048C
Trigger sensitivity Calibrator 25 MHz, 60 MHz and 100 MHz sine
waves.
FLUKE 5520A
Power splitter Outputs differ < 0.15 dB Agilent 11667B
Cable BNC, quantity of 3 Agilent 10503A
Adapter N (m) to BNC (f), quantity of 3 Agilent 1250-0780
Feedthrough 50&, BNC connector (m) and (f),
quantity of 2
Agilent 11048C

Table of Contents

Other manuals for Agilent Technologies U1604B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1604B and is the answer not in the manual?

Agilent Technologies U1604B Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1604B
CategoryTest Equipment
LanguageEnglish

Related product manuals