Temperature measurement problems and recommended actions (continued)Table 10-3:
Problem Possible causes Recommended actions
Inaccurate tempera-
ture data from exter-
nal device
Wiring problem
Problem with input configuration
Problem with external device
Problem with input configuration
Verify the wiring between the transmitter
and the external device.
Verify that the external device is operating
correctly.
Verify the configuration of the tempera-
ture input.
Ensure that both devices are using the
same measurement unit.
10.9 API referral problems
Problem Possible causes Recommended actions
No referred density
reading
Line pressure, line temperature, or line
density is outside the range of the config-
ured API table
Check process conditions.
Ensure that the API referral application is
configured correctly for your process fluid.
See Section 4.5.
Inaccurate referred
density reading
Inaccurate density measurement
Inaccurate temperature measurement
Incorrect reference conditions
Incorrect API table selection
Verify the line density value. If it is not ac-
curate, see Section 10.7.
Verify the line temperature value. If it is
not accurate, see Section 10.8.
Ensure that the application is configured to
use the appropriate temperature source.
Ensure that reference temperature and ref-
erence pressure, if applicable, are config-
ured correctly.
Ensure that the selected API table is appro-
priate for the process fluid.
10.10 Concentration measurement problems
Problem Possible causes Recommended actions
Significantly incorrect
concentration meas-
urement after loading
matrix
The wrong temperature or density unit
was configured when the matrix was loa-
ded
Set the temperature and density units to
the units used when the matrix was built,
then reload the matrix. For custom matri-
ces, contact Micro Motion.
Troubleshooting
148 Micro Motion
®
Fork Density Meters (FDM)

Table of Contents

Other manuals for Emerson Micro Motion

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Micro Motion and is the answer not in the manual?

Emerson Micro Motion Specifications

General IconGeneral
BrandEmerson
ModelMicro Motion
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals