2.6 Safety and Device Protecting Features
The device is equipped with following safety and device protecting features :
- High current protection which prevents damage to the device and potential danger to
person, e.g. in case of short circuit.
2.7 Measurement Accuracy Verification
Maintenance: The ergometrics 900L ergometer must be regularly inspected and main-
tained, in time intervals of two years, in accordance with lawful regulations. Alterations
to the ergometer’s measurement capabilities are cause for re-certification.
2.7.1 Routine Control of the Blood Pressure Measurement Unit
The automatic blood pressure measurement unit in the console of the ergometer must
be approved and recalibrated every two years by personnel authorized by ergoline.
2.7.2 Periodic Accuracy Verification for Blood Pressure Unit
EC-Directive 93/42/EEC operator regulations stipulate that the automatic blood pres-
sure unit in the ergometer´s console must undergo a routine technical inspection every
two years. (see page II in this operators Manual)
2.7.3 Load Unit Inspection
EC-Directive 93/42/EEC operator regulations requires qualified technical maintenance
of the load unit at two years intervals. This is to be performed by an ergoline authorized
service technician.