6.2.2 Entries Peculiar to the use of Program 0
When using program 0, the maximum heart rate (HR) should always be entered spe-
cific to the individual patient. Since this value serves as the target frequency for load-
ing, it must be specifically adapted to the age and health of the patient. The remaining
parameters, in particular the alarm parameter, should be established in the configura-
tion so that even the weak patient can withstand the ergometric session without danger.
Here also, all other parameters may be altered for a specific ergometric session.
!
Always enter the maximum heart rate (HR) for program 0 specific to the
age and strength of the patient. When possible, provide the ergometer
with a continuous heart rate by using an ear/finger clip, a pulse monitor or
the HR feedback from an ECG
6.2.3 Entries Peculiar to the use of Program 9
Patient specific maximum heart rate (HR) should also be entered for program 9 to
reflect the patient’s training requirements.
!
Only use program 9 when a continuous heart rate source is used. That is,
when an ear/finger clip, a pulse monitor or a HR feedback from an ECG is
connected.
Ergometry process