Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
3 - 50 Section 3-7 - Setting Up Connectivity and Tips
3-7-4-4 DICOM Export
Use DICOM Export to export DICOM content in either .mp4 and .avi formats.
Table 3-17 To Configure DICOM Export
Step Corresponding Graphic
1.
Navigate to Connectivity > Connection Manager > MyComputer
Device.
2.
Select DICOM Export > Configure.
3.
In Name, enter a logical name of the service.
4.
To have raw cine clips converted to DICOM multi-frames during EzBackup operation, enable Convert Clips.
5.
To have raw volumes get converted to Enhanced US Volume (DICOM 3D) format during EzBackup operation, enable Convert Volumes.
6.
To round measurement values to configured number of decimal places, enable Apply Display Precision to Measurements.
7.
Under Dicom Coding Scheme, select how measurements are encoded with configured coding scheme in DICOM SR objects.
8.
To use button settings, enable Use Button Settings.
9.
Select Done.