Chapter 5: Quick Steps
5495975-1EN Rev.9 5-4
© 2013-2017 General Electric Company. All rights reserved.
CAUTION Potential Pinch Point: The area where the tube connects to the column may create
a pinch point when the tube is rotated. Operators should keep their hands on the
OTS handle and keep patient’s clear while rotating the tube.
Rotate the Multi-Leaf Collimator
1. Move the locking lever on the multi-leaf collimator toward the front panel, i.e. toward you. This
releases the collimator from the 0 lock-in position.
2. Grasp the multi-leaf collimator with both hands.
3. Rotate it to the desired angle and direction.
4. Return the locking lever to its original position. This prevents further rotation.
Note: When the collimator is rotated, the image border is adjusted to the maximum size based on the
selected collimator FOV.
CAUTION Always grasp the multi-leaf collimator in such a way that neither hand can be
pinched or crushed between the handles and the collimator.
General Acquisition
This section outlines the basic acquisition process.
Refer to Chapter 9: Worklist
for more detailed information about adding or selecting procedures to per-
form.
Refer to Chapter 10: Image Acquisition
for more detailed information about the Acquisition screen func-
tions, how to conduct specific types of exams, and how to end exams.
1. Add or select the procedure from the Worklist.
CAUTION Select all procedures for the patient.
2. Click [START EXAM] or [RESUME EXAM].
3. Select the protocols for the exam.
4. Click [ACCEPT].
5. If multiple procedures were selected from the Worklist, choose the exam to perform.
6. Choose the protocol to perform.