Chapter 13: Advanced Applications
5495975-1EN Rev.9 13-9
© 2013-2017 General Electric Company. All rights reserved.
6. Configure the preset names for the Operators, if necessary. Refer to Chapter 15: Preferences-Preset
Names (p. 15-19).
RRA Confirmation Screen
The RRA Confirmation Screen selection controls the presence of the confirmation screen displayed on
the viewer when the exam is closed. This screen provides a summary of the RRA classified images in the
exam.
Follow this process to enable or disable RRA Confirmation Screen:
1. From the Worklist screen, click [UTILITIES].
2. Select Preferences > Repeat/Reject Analysis.
3. Click RRA Confirmation Screen [ON] to enable the function.
ï‚· RRA Confirmation Screen [OFF] disables the function.
4. Click [CLOSE].
Export RRA Data and Report
The system stores RRA data for the last 100 days. This data may be exported in a report to a disk or USB
device.
IMPORTANT! RRA data is not saved with system backups. RRA data should be exported prior to per-
forming any software updates to the system.
IMPORTANT! RRA Data cannot be exported on DVD. Only CDs and USB devices can be used for RRA
export.
The exported data includes a summary report in html format (Figure 13-6) and a Microsoft excel spread-
sheet (Figure 13-7) with acquisition and classification details for all repeated and rejected images on the
system. The spreadsheet also contains JPEG links for all repeated and rejected images on the system.
The Excel spreadsheet includes the following information for each repeated/rejected image:
IMPORTANT! The RRA export data does not contain patient names: however, patient ID numbers, image
JPEGs, and accession number data are present. Proper HIPAA or privacy and internal or
local confidentiality standards should be practiced.
ï‚· System ID
ï‚· Automatic Protocol Code
ï‚· Acquisition Date
ï‚· Acquisition Time
ï‚· Processing Date
ï‚· Processing Time
ï‚· RRA Classification
ï‚· Repeat Reason
ï‚· Operator Name
ï‚· Anatomy
ï‚· View
ï‚· Acquisition Type
ï‚· Link to JPEG Image
ï‚· Patient ID
ï‚· Accession Number
ï‚· Size Selection
ï‚· kVp
ï‚· mAs
ï‚· mA
ï‚· SID (mm)
ï‚· Grid Status
ï‚· Filtration (mm Cu)
ï‚· Manual/AEC
ï‚· Ion Chambers Selected
ï‚· DAP
ï‚· DEI
ï‚· DEI Limits