GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 3 - Setup Instructions 3-55
Section 3-8
System Configuration
3-8-1 System Setup
Modifications of system parameters and settings are supported by 7 major groups.
Each maj group contains diverse dialog pages and sub windows.
• General Settings: Date, Time, Clinic Name, (EUM) Language, Screen saver, etc.
• Details: Doppler 2D Refresh, Zoom options, Menu Brghtness, Trackball Speed, etc.
• Annotation: adjustment of different Annotation presets(e.g. Home position, etc.)
• Clipboard: adjustment of Clipboard display and functions
• Service tab: conncet the USB “standard GE” dongle and enter the “standard GE revolving”
password (change every six months_ to get access to the Service Tools functions.
• System Info: shows which Software/Hardware version is installed in the system
• Options: shows which options are installed in the system
For information on configuring Software Options refer to Section 8-7 on page 8-23
• Peripherals: Video Norm selection, Foot switch assignment, Add Printer, Edit Printer settings, etc.
• Device Setup: to set up all DICOM, Archive and Network configuration nodes
• Button Configuration: to adjust assignment of Remote keys P1
, P2 and P3... (e.g, Printer
selection)
• Drives: USB and Network drives: stop devices, map network drive, Erase CD
• System Configuration: Save/Load Settings only, Save/Load/Delete Full System Configuration
• Image Archive: to save or load Image Archive
• Presets: to save User Folder/Presets and 3D/4D Presets, Logo display, etc.
• Global Parameters: adjustment of different parameters for specific or all applications
• Menu to program Biopsy Lines
• Measure & Calc: Shows all settings, which are used for generic measurements as well as
calculations in different applications
• Application Parameters: to adjust Status on Freeze for different modes, Manual Trace method,
Calculation Ratio, etc.
• Global Parameters: to select cursor type and size, font size and color of measure results position
of results for different modes, etc.
Report: adjustment and setup of Report display
Result Windor: adjustment of Result and OB Graph display
Measure Button: adjust assignment of Measure buttons
NOTICE
!! NOTICE:
More detailed information pertaining System Setup adjustments is found in the Basic User Manual; see:
Table 9-10, ?System Manuals - Voluson™ P8 / Voluson™P6,? on page 9-19.

Table of Contents

Other manuals for GE Voluson P6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P6 and is the answer not in the manual?

GE Voluson P6 Specifications

General IconGeneral
BrandGE
ModelVoluson P6
CategoryMedical Equipment
LanguageEnglish

Related product manuals