GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 3 - Setup Instructions 3-15
3-5-3 Connecting the Color Printer
1.) Power OFF/Shutdown the system as described in: Section 3-6-3 on page 3-28.
2.) Connect the Color printer according to correct connection scheme as described in Figure 3-7 on
page 3-16.
NOTE: The Color printer should be connected to the USB1 port of the Voluson™ P8/Voluson™P6´s PC-part.
3.) When all the cables are connected, press the Power ON switch on the Color printer.
4.) Power ON/Boot up the Voluson™ P8/Voluson™P6 system as described in Section 3-6-2 on page
3-25. All software drivers are pre-installed for the designated Color printer only.
5.) After physical connection to the Voluson™ P8/Voluson™P6 system, assign the printer to a remote
key (P1
, P2, and P3 ) as described in Section 3-7-8 "Remote Control Selection" on page 3-53.
6.) Verify the correct settings in the printer “Properties”, see: Section 3-7-7 "Adjustment of Printer
Settings" on page 3-49.
WARNINGWARNING
!! WARNING:
After each installation, the leakage currents have to be measured according to
IEC 60601-1 respectively UL 60601-1 and ANSI/AAMI ES60601-1.

Table of Contents

Other manuals for GE Voluson P6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P6 and is the answer not in the manual?

GE Voluson P6 Specifications

General IconGeneral
BrandGE
ModelVoluson P6
CategoryMedical Equipment
LanguageEnglish

Related product manuals