GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-137
Section 8-45
Replacement of the DBM64G
8-45-1 Manpower
One person, 30 minutes
8-45-2 Tools
Philips screwdriver
8-45-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
2.) Remove the REAR CABINET BEZEL ASSY BT16 and disconnect cables as described in 8-28-4
"Removal Procedure" 2), 3), 4) and 7) on page 8-93.
8-45-4 Removal Procedure
1.) Pull the DBM64G/DPI out of the Nest Frame
Figure 8-196 DBM64G/DPI
Figure 8-197 Drawing the DBM64G/DPI
DBM64G