GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-9
Section 7-3
Check Points Voltages
7-3-1 How to check power
7-3-1-1 Dachstein Power Supply (DPS)
Turn on mains switch and check green LED of P5V, N5V, LIVE5V and 12V on backplane.
NOTE: If LED is on, DPS has full function (all voltages are within the specified range).
Figure 7-6 Check green LED on backplane