GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
8-126 Section 8-40 - Replacement of the SPEAKER COVER LEFT/RIGHT
Section 8-40
Replacement of the SPEAKER COVER LEFT/RIGHT
8-40-1 Manpower
One person, 5 minutes
8-40-2 Tools
Philips screwdriver
8-40-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
8-40-4 Removal Procedure
1.) Remove 2 RUBBER CAPS on SPEAKER COVER LEFT/RIGHT. Refer to the figure below.
2.) Unscrew 2 screws, then remove the SPEAKER COVER LEFT/RIGHT. Refer to the figure below.
8-40-5 Installation Procedure
1.) Install the new parts in the reverse order of removal.
Figure 8-184 Removing 2 RUBBER CAPS on SPEAKER COVER LEFT/RIGHT
Figure 8-185 Unscrewing 2 screws to remove SPEAKER COVER LEFT/RIGHT