GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-91
Section 8-26
Replacement of the SPEAKER
8-26-1 Manpower
One person, 30 minutes
8-26-2 Tools
Philips screwdriver
8-26-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
8-26-4 Removal Procedure
1.) Remove the MAIN OPIO and SUB OPIO. Refer to the Section 8-25 on page 8-87 OPIO.
2.) Disconnect the 2 SPEAKER CONNECTORS. Refer to the figure below.
3.) Unscrew each 3 screws of SPEAKER ASSY L, R and remove. Refer to the figure below.
8-26-5 Installation Procedure
1.) Install the new parts in the reverse order of removal.
2.) Perform: FRU8-11: Replacement of the SPEAKER - Functional Tests.
Figure 8-136 Disconnecting SPEAKER CONNECTOR
Figure 8-137 Unscrewing 6 screws to separate SPEAKER ASSY